Back to Journals » Clinical Ophthalmology » Volume 18

Current Concepts and Recent Updates of Optical Biometry- A Comprehensive Review

Authors Pathak M, Sahu V , Kumar A, Kaur K, Gurnani B 

Received 17 February 2024

Accepted for publication 28 April 2024

Published 2 May 2024 Volume 2024:18 Pages 1191—1206

DOI https://doi.org/10.2147/OPTH.S464538

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2

Editor who approved publication: Dr Scott Fraser



Madhumallika Pathak,1 Vijaya Sahu,2 Aseem Kumar,2 Kirandeep Kaur,3 Bharat Gurnani4

1Department of Ophthalmology, Sai Baba Eye Hospital, Raipur, Chhattisgarh, 492099, India; 2Department of Ophthalmology, All India Institute of Medical Sciences, Raipur, Chhattisgarh, 492099, India; 3Department of Pediatric Ophthalmology and Strabismus, Gomabai Netralaya and Research Centre, Neemuch, Madhya Pradesh, 458441, India; 4Department of Cataract, Cornea and Refractive Surgery, Gomabai Netralaya and Research Centre, Neemuch, Madhya Pradesh, 458441, India

Correspondence: Vijaya Sahu, All India Institute of Medical Science, Raipur, Chhattisgarh, 492009, India, Tel +91975267955, Email [email protected]

Abstract: One of the most recent advancements in the field of cataract surgery is optical biometry. With the advent of optical biometry ocular measurements are now simpler, quicker, and more precise. The devices have made intraocular lens (IOL) power calculations easier in difficult situations too, such as in cases with extremes of axial lengths, silicone filled eyes, cataract surgery in post-keratoplasty eyes, post Laser-Assisted in Situ Keratomileusis (LASIK) eyes, etc. The gold standard for IOL power calculation in the present day is by the use of optical biometry devices. The anatomical measurements by these devices are highly precise and because of these measurements and the incorporation of various IOL power calculation formulas the optical biometry devices give the accurate power and the post-operative visual outcome is highly satisfactory among the patients. The growing use of these devices has made cataract the most commonly performed refractive surgical procedure nowadays. In the current scenario, optical biometry has widespread acceptance in almost all countries and has many advantages over ultrasound or immersion biometry. Cataract surgeons can obtain easy and reliable measurements from these devices. Refractive surprises have also decreased considerably with their use. This article will comprehensively review the principles of the various optical biometry devices, the parameters used in each of the devices, the advantages and disadvantages, and add more like what all this article will add.

Keywords: optical biometry, IOL master 500, IOL master 700, lenstar LS 900, OA 2000, aladdin

Introduction

Biometry is the practice of applying mathematics to biology, and it is utilized in preoperative measurement for intra-ocular lens (IOL) power calculation before cataract surgery.1 Evaluating myopia progression through axial length measurement is crucial because axial elongation is a definitive indicator of myopia development and progression. Monitoring changes in axial length can help predict the rate at which myopia is worsening, which is especially important in pediatric populations where interventions may be most effective.2 Additionally, analyzing the cornea using topography and tomography is vital in identifying and monitoring corneal abnormalities. Corneal topography provides detailed maps of the corneal surface curvature, essential for diagnosing and managing conditions like keratoconus.3 Tomography offers a three-dimensional assessment, giving insights into the corneal structure that are crucial for a variety of clinical applications, including refractive surgery planning and the diagnosis of corneal diseases. Together, these biometric evaluations play a significant role in comprehensive eye care.4 Cataract remains the primary cause of visual impairment globally, and accordingly, cataract extraction is the most frequently performed operation in the field of ophthalmology.5 Currently, the term “refractive cataract surgery” has emerged where the goal of a cataract surgeon is not just to eliminate the blindness or defective vision caused by the cataractous lens but also to provide crystal clear spectacle-free vision postoperatively. The factors that play to give patients customized, clear, and spectacle-free vision are the surgical procedures used like topical phacoemulsification, femtosecond laser-assisted cataract surgery (FLACS), implantation of various types of premium IOLs with the precise and accurate biometry. The highly advanced optical biometry devices, with the incorporation of new IOL power calculation formulas, make the ocular measurements very precise and can provide accurate IOL power to be implanted intraocularly.6,7

There are several biometric systems for the measurement of different ocular structures that actually determine the precision in IOL power calculation. These systems are Ultrasound A Scan (USG-A) and Ultrasound B-scan, low coherence interferometry, and laser interferometry.8 These systems are used for the measurement of axial length, which is the most important variable in IOL power calculation. Other measurements are keratometry (K), anterior chamber depth (ACD), lens thickness (LT), pachymetry, and retinal thickness (RT). Axial length (AL) and K are the deciding factors and any preoperative measurement errors of either of these alter the post-operative refraction and are responsible for unpleasant refractive surprises. According to some studies, 54% of the fallacies in IOL power calculation are due to error in AL measurement, 38% due to incorrect estimation of post-operative ACD, and 8% because of corneal power measurement error.9 Hence, the role of optical biometry devices comes into play which have higher precision and better accuracy and have incorporated newer IOL power calculation formulas. So, we can say that the outcome of refractive cataract surgery depends on these various modern state-of-the-art devices.

Until the 1970s, Ultrasound Biometry was the benchmark for axial length measurement. In 1999, the initial optical biometer made commercially available was the IOL Master 500 (Carl Zeiss Meditec AG in Jena, Germany). This device operates on the foundational principle of partial coherence interferometry. It has 8 times more resolution than 10 MHz sound wave, used in ultrasound biometry by utilization of 780-micron infrared light wave. There are some differences between Ultrasound biometer and optical biometry, which are highlighted in Table 1.10–12 Although there are many advantages of optical biometry like noncontact method, non-invasive method, and very precise technique, there are certain drawbacks of optical biometry in comparison to US biometers, the cost of optical biometers is quite high and measurements are not possible with optical biometry in patients who are not able to fixate eyes, like in nystagmus and squint. Also, some difficulty while measuring the AL in patients with media opacities, e.g., corneal opacity, or dense cataracts.10–12

Table 1 Difference Between Ultrasound and Optical Biometry

The different optical biometry devices work on different principles. Many researches have come up regarding the newer and more advanced devices with their merits, but there needs to be more literature on a comprehensive review. In this article, we have reviewed the principles as well as various advantages of the recent advances.

Literature Search

We did an electronic search in PubMed for articles published from 1990 to 2022. We searched with the terms ‘Optical biometer’, ‘principle of optical biometer’, ‘ultrasound biometry’, ‘advanced optical biometer’, and ‘cataract refractive surgery and optical biometers’. We excluded the non-English articles. We have cited the most relevant and recent indexed articles.

Principles of Optical Biometry

Partial Coherence Interferometry was first time used for AL measurement in 1986 by Fercher et al.13 Later, Zeiss used this technology and launched the first commercially available optical biometer- IOL Master 500, in 1999.14 Since their launch, several improvements have been made, especially for dense cataracts. Newer Optical Biometry devices could cut down on background noise through dense cataracts. The current optical biometers use either optical low coherence reflectometry (OLCR) or swept-source optical coherence tomography (SS-OCT). Several newer devices are also available now. Figure 1 and Table 2 show the different types of Optical Biometers15–22 In this article, we have highlighted the principles, advantages, limitations, and current literature on some of the Optical biometers that are commonly used in India and try to give an overview of recently launched devices.

Table 2 Depicts the Various Type of Optical Biometers Available Along with the Principle Used

Figure 1 Flow chart showing different types of Biometers.

Partial Coherence Interferometry Optical Biometer

IOL Master 500

The IOLMaster, introduced in 1999, marked the beginning of a new era in intraocular lens (IOL) measurements, and the technology saw its next major advancement in 2010 with the introduction of the IOLMaster 500.23

Principle

IOL Master 500 (Carl Zeiss Meditec AG, Jena, Germany) is based on the principle of Partial Coherence Interferometry (PCI).24 A semiconductor laser diode emits a dual beam of infrared (IR) light (780 nm). A signal is produced as a result of interference between the light reflected from the tear film and that reflected by the retinal pigment epithelium. The interference signal is received by the photodetector which is used to calculate the optical distance (OD) between the corneal surface and the retina. This OD is used to attain the other geometrical intraocular distances. The parameters measured by IOL Master 500 are axial length (AL), keratometry (K), anterior chamber depth (ACD), and white-to-white distance (WTW). At a 2.5 mm zone on the anterior cornea, six measured points calculate K. ACD is measured with the help of lateral slit illumination.25 The IOL power calculation formulas included are Holladay I, Holladay II, SRK II, SRK/T, Haigis, and Hoffer Q. The device, incorporates Haigis formula calibration, which is based on immersion ultrasound biometry. This calibration process is essential for the precise calculation of intraocular lens (IOL) power required for cataract surgery.26 The Haigis formula, one of the many biometric algorithms available for IOL power prediction, uses AL along with ACD and LT measurements, to provide highly accurate IOL power estimations that can enhance postoperative visual outcomes. It also utilizes a Group Refractive Index (GRI) to measure axial length (AL) for the calculation of intraocular lens power. The GRI helps in accounting for the varying refractive indices of the different media that the light passes through within the eye.27,28

Advantages

  • The system delivers keratometry readings that are not influenced by distance, ensuring consistent and replicable measurements. It exhibits remarkable congruence with traditional manual keratometry, yet surpasses it in terms of measurement precision.29
  • Provided accurate markerless toric IOL alignment
  • The ZEISS IOL Master 500 demonstrates a notably expedited measurement process, capable of capturing readings approximately four times more swiftly than alternative optical instruments. It has the efficiency to assess both eyes within a period of under one minute.30

Limitation

By IOL Master 500, measurements are practically difficult through corneal opacities and dense cataracts.

Optical Low Coherence Reflectometry Ocular Biometer

Lenstar LS 900

Lenstar-LS900 (Lenstar) (Haag-Streit AG, Koeniz, Switzerland) was introduced in 2008 based on OLCR technology.31 The ocular measurements AL, ACD, lens thickness (LT), central corneal thickness (CCT), and retinal thickness are taken using an 820 nm super-luminescent diode.30 In addition, it is also based on GRI.32

Advantages of Lenstar Over IOL Master 500

  • More accurate measurement of ACD – as it measures the aqueous depth which is measured from corneal endothelium rather than from anterior corneal surface by IOL Master 500.33
  • Due to the dual-zone analysis of the light-emitting diode projection at 1.65 and 2.3 mm of closely spaced 32 measurement points, the K reading calculation is more precise.
  • Lenstar measures the LT, RT, size, and centricity of the pupil; these are not available in IOL Master 500.
  • It provides more accurate biometry results because of the incorporation of the latest IOL power calculation formulas (Barrett, Olsen, Holladay 2).
  • The Hill-radial basis activation function (Hill-RBF), Barrett Universal II, Barrett True-K, and Barrett Toric calculator are included in the most recent version of Lenstar.34

The Lenstar LS 900 non-contact biometer’s reproducibility was assessed by Cruysberg et al30 on 76 eyes of 38 healthy volunteers. The reproducibility of the Lenstar was found to be outstanding when compared to the Visante anterior-segment optical coherence tomography (AS-OCT) and the IOLMaster. No clinically significant difference was observed in the IOL power computation findings, despite the three devices’ AL readings differing significantly from one another

Limitations of Lenstar

  • The AL measurement range is less compared to the IOL Master 500, that is 14–32 mm for the Lenstar compared to 14–40 mm by IOL Master 500.
  • Biometry is difficult to perform through dense cataracts.34

AL Scan (Nidek)

In 2012, Nidek introduced the AL-Scan optical biometer to the market, expanding its portfolio of ophthalmic diagnostic equipment.35 The principle of optical low-coherence interferometry is used in the Nidek AL-Scan (Nidek CO., Gamagori, Japan) and it measures K values utilizing double-mire rings projected onto the cornea at the 2.4 mm and 3.3 mm zones. The light-emitting diode performs WTW assessment and corneal keratometry readings.15,36,37 The distance between light reflections on the corneal and lens anterior surface is used to assess ACD. An anterior eye segment eye image is used to calculate the WTW.36,37 Huang J et al36 evaluated measurements by AL scan, compared them with IOLMaster 500 and concluded that with the exception of WTW and pupil distance, AL-Scan’s repeatability and reproducibility were outstanding. There was good agreement between the AL-Scan and IOLMaster, except WTW.

Swept-Source Optical Biometer (SS-OCT)

IOL Master 700

In 2007, IOL Master 500 was upgraded to version 5, that have made significant changes to its technology.

Carl Zeiss Meditec launched the IOLMaster 700 in 2015, and by May 2017, it was accessible in major markets, including the United States.38 Thereby measurements have become possible even through opaque media (corneal opacity, dense cataract) and there is an increase in the accuracy of axial length and keratometry readings.39

Principle of IOL Master 700

The first biometric device based on swept-source optical coherence tomography (OCT) is the IOL Master 700 (Carl Zeiss Meditec AG Jena, Germany).40 It enables OCT imaging and viewing along the entire eye’s length. A 1055nm wavelength laser source is used to scan the eye. The longitudinal section of the entire eye can be viewed as an image-based measurement is provided. SS-OCT based biometers apply an optical B scan to get biometric data. It measures AL, K (2.5 mm zone), CCT, ACD, LT, WTW corneal diameter, and pupil diameter.41

Advantages

  • It is able to detect aberrant eye geometries, including lens tilt.
  • Insufficient fixation can also be identified as it images the fovea.
  • For corneal power measurements, it also uses telecentric keratometry similar to the IOL Master
  • Swept-source OCT gives us total keratometry (TK), i.e., measures the posterior corneal surface. Total Keratometry can be used in classic IOL calculation formulas, and there is no need for any additional software or an online calculator. Furthermore, Barrett TK Universal II, Barrett TK Toric, and Barrett True K with TK are the three formulas that Graham Barrett has created specifically for Total Keratometry. In post-myopic LASIK eyes, Barrett True K formula with TK enhanced the result prediction compared to the Barrett True K with Classic Ks within ±0.5 D by >12% (p = 0.04)41
  • There is no need to use a separate online toric calculator as IOL Master 700 contains an inbuilt toric calculator (Barrett Toric calculator and Haigis‑T for toric IOLs).40

In their study, Akman et al40 evaluated and compared the new swept source OCT-based IOL Master 700 with the IOL Master 500 and concluded that in biometric measurements in eyes with posterior subcapsular and dense nuclear cataracts, IOL Master 700 was more effective. Recently IOL Master 700 has come up with software Update 1.90 with Central Topography. The latest software upgrade enhances corneal contour assessments, enables surgeons to examine surgical planning details on their mobile devices via ZEISS’s EQ Mobile application, and includes the advanced Barrett True K with Total Keratometry calculation.

Limitation

There are no major limitations with IOL Master 700, it is considered one of the best optical biometry devices. It is user and patient-friendly, highly accurate, and widely used in almost all the institutes at present. However, to list a few:

  • The IOL Master is based on Group Refractive Index (GRI). GRI-based biometers gave longer AL measurements in long eyes and shorter ALs in short eyes compared to sum-of-segments biometers.
  • In addition, GRI-based biometers reliability could be affected by lens opacity.42

Argos Advanced Optical Biometer

The ARGOS Advanced Optical Biometer made its debut in the year 2015.43 The Argos Advanced Optical Biometer is based on coherent optical interferometry and tomography with lateral scanning of a 1-μm swept-source beam. A ring of 16 infrared LEDs provides illumination, which produces keratometry (K) readings. The corneal curvature data is produced by combining the OCT signal with the reflected image from the LEDs. Parameters measured include AL, CCT, ACD, LT, K values, pupil size, and toric axis.44,45

Advantages

  • In comparison to other partial coherence interferometry devices, the sensitivity of the Argos is 10 times greater.
  • In Dense Cataract mode, it can increase sensitivity up to 100 times for AL detection, resulting in improved success rates.
  • In comparison to conventional biometry, ARGOS exhibits faster and more accurate biometry. It also offers a unique live 2D OCT view of the entire eye, from cornea to retina and limbus to limbus.
  • One of the greatest advantage of ARGOS is, it can successfully measure the axial length in denser cataracts. By utilizing Swept-Source Optical Coherence Tomography, or SS-OCT, it ensures correct ocular biometry for the selection of IOLs by using SS-OCT – Swept-Source Optical Coherence Tomography.44,45
  • It does not work with GRI but it used single refractive index for each ocular structure (cornea, aqueous, lens, vitreous), obtaining the so called ”sum of segments”.42

OLCR with Placido Disc Corneal Topography Optical Biometer

Aladdin

The Aladdin Optical Biometer with OLCR and Placido Disc Corneal Topography was first introduced before July 2015.46 Corneal topography and an optical biometer are combined in the Aladdin (Topcon) device. This device operates on the basis of optical low coherence interferometry (OLCI), which uses an 830 nm super-luminescent diode to measure the AL of the eye. Placido ring topographer creates a corneal topography utilizing the reflection of 24 numbers of 8 mm-diameter Placido disk rings.16,36

Advantages

  • The data for automated keratometry are produced by reflecting four specific Placido rings, totaling 1024 points, with a diameter varying from 2.4 mm to 3.4 mm, so additional information of corneal asphericity is obtained by Aladdin.
  • Because of the incorporation of corneal topography, any type of corneal irregularities like incipient keratoconus, and higher-order aberrations can be detected, which can help surgeons in the selection of the IOLs.
  • Along with corneal topography, it also provides the assessment of pupillometry, which is becoming increasingly popular for the selection of premium IOLs.
  • It has Barrett IOL calculation formula suit incorporated, which is helpful not only in the selection of Toric IOL but also in any type of IOLs and neither any AL adjustment is required.

Limitations

Some studies found differences in K and ACD measurement compared to other optical biometers.13,32 Kenneth J Hoffer et al47 conducted a multicentric study wherein measurements provided by Aladdin were compared with those provided by the IOL master 500 and concluded that no statistically significant difference was found in AL values. However, Aladdin gave lower mean Keratometry values and deeper ACD measurements, that needed constant optimization when calculating the intraocular lens power using theoretical formulas. Huang J et al36 assessed the precision of Aladdin in patients with cataracts and in healthy subjects. In a total of 98 people enrolled, 46 eyes were from patients with cataracts, and 52 eyes were from healthy subjects and concluded that Aladdin demonstrated excellent intraoperator repeatability and interoperator reproducibility for AL, ACD, and K values measurements in both groups. However, in patients with cataracts the precision of WTW measurements was lower.

OA 2000 (Tomey)

In 2014, the OA‑2000 was launched by Tomey (GmbH, Nurnberg, Germany).48 It measures ocular biometry using the low coherence reflectometry (OLCR) technique. This device measures the pupil size, CCT, WTW diameter, LT, AL, and K values. [51,52} The corneal topography is measured using a Placido disc-based topography, which projects nine rings onto the cornea, each with 256 points, within a 5.5 mm zone. Here, high-speed tissue penetration using the Fourier domain technology is combined with an autonomous search mechanism for measurements of CCT, ACD, LT, pupil diameter, WTW diameter, and AL.

Advantages

  • Even with dense cataracts, a search algorithm may automatically detect measurements.
  • It does not require realignment and can execute 10 consecutive scans for each measurement.
  • The AL-4000 handheld ultrasound device, which pairs via Bluetooth with the OA-2000, can be utilized for mature cataracts.
  • The OA-2000 is lightweight, quick, easy to use, and patient-friendly.49,50

Limitations

  • Not good for WTW diameter.
  • Its accuracy has so far been evaluated only in few studies.

Kongsap compared the new optical biometer (OA 2000) and a standard biometer in his study51 on 102 eyes of 68 cataract patients. He found that for nearly all ocular biometry measurements, the OLCR biometer showed a very strong agreement with the standard PCI optical biometer, with the exception of the WTW diameter. Anterior and posterior segment OCT is not incorporated

Revo NX 130

In 2017, OPTOPOL Technology introduced significant devices to the market: the REVO NX, which was recognized as the fastest OCT at that time.51 One more recently introduced optical biometry is OCT Biometry (B-OCT) that does the ocular axial dimensions measurement using a conventional OCT system. By altering software and technique of ocular scan of a commercially available OCT device, B-OCT was implemented, and that is the newer device Revo NX.50,52 Version 9.0 of the Revo NX software (Optopol Technology Ltd, Zawiercie, Poland) is a fast spectral-domain OCT with110000A-scan/sec speed. It features an add-on lens that allows it to measure the axial length, see the posterior segment of the eye, and make maps of the cornea and anterior segment images. Within an 8 mm corneal diameter, 16 B-scans can be automatically obtained. Anterior, posterior, true corneal power, and CCT are assessed from the center 3 mm zone.

Advantage

A single machine that incorporated OCT for the anterior and posterior segment and optical biometer.

Limitation

Very dense nuclear cataracts make biometric assessments challenging.

Initially introduced by Bartosz L. Sikorski et al,50 OCT biometry (B-OCT) is a novel method for assessing ocular axial dimensions. In 349 eyes examined, (214 healthy individuals, 115 cataract patients, and 20 eyes with severe macular disorders), B-OCT was used for the first time in the spectral domain OCT equipment for posterior and anterior segment imaging (REVO NX, Optopol Technology). Following a comparison of the B-OCT results with the Carl Zeiss Meditec swept source OCT-based IOL Master 700, they came to the conclusion that very small and nonsignificant differences were found between the biometric measurements recorded using REVO NX B-OCT and IOL Master 700. Both intra-observer and interobserver reproducibility showed great precision for B-OCT.50

OLCR with Rotating Scheimpflug Camera

Galilei G6

The Galilei G6, a cutting-edge device designed for detailed corneal and ocular analysis, received its FDA approval in July 2019.53 The Galilei dual Scheimpflug analyzer (Ziemer, Switzerland) combines two rotating Scheimpflug cameras with a Placido disk to image the anterior segment of the eye. Galilei provides two- and three-dimensional anterior segment imaging, lens densitometry, anterior and posterior corneal topography, and comprehensive corneal pachymetry. Axial biometry is carried out by low coherence interferometry using light with a wavelength of 880 nm. The intraocular lens power calculation is produced by combining the biometric measurement and anterior segment measurements with the Galilei G6.20,54

Advantage

Detailed assessment of the cornea, i.e., presence of keratoconus, and any high-order aberrations prior to cataract surgery, which in turn helps in IOL selection.

Jung S et al in their study in 101 eyes of 54 patients, compared the repeatability and agreement between the IOL master 700 and Galilei G6 and concluded that both the biometers showed high repeatability and relatively good agreements. However, because it uses a 1055 nm tunable laser source, which can enter tissue more effectively and with less scatter, the swept-source optical biometer (IOL master 700) showed superior repeatability, penetration, and an overall reduced prediction error. Additionally, the arc scan pattern used by the SS-OCT optical biometer for biometric measures may enhance the device’s penetration potential.20

Recent Advances

The optical biometers have brought tremendous changes in IOL power calculation, and now the surgeons can customize the type of IOLs according to the need of the patients. These devices have eased the calculation in every situation, right from post-traumatic to post-refractive surgery eyes. However, calculation in denser cataracts with some of the devices is still difficult. To overcome this, new devices with newer technologies have been developed. Additionally, not only a correct biometry, but also a good choice of the right IOL power calculation formula designed for post-refractive surgery eyes is essential.55 A recent review has documented that measurements of axial length in dense cataracts are successful with newer swept-source ocular tomography biometers and precision in toric IOL placement is improved. Additionally, it is not required to have patients take out their soft contact lenses longer than two days before biometry.56 Another recent review study came to the conclusion that the gold standard for determining axial length in cataracts of any kind is SS-OCT. We have mentioned a review of literature of a few of the recent clinical studies on various optical biometers in Table 3.

Table 3 Review of Literature of Various Studies Employing Using Optical Biometers

Conclusion

The Biometry devices, from Ultrasound to Optical Biometry have evolved over the years tremendously using various advanced technologies. For many years, scan ultrasound was thought to be the best option available, but today we have excellent imaging systems that have incorporated many newer generation IOL power calculation formulas by which 90–92% emmetropia can be achieved in post-operative patients.71 The use of these modern devices has highly increased patients’ satisfaction in terms of clarity and quality of vision. Looking at the increasing accuracy of more than 90% of the technologically advanced devices, it is undoubtedly certain that the future of biometry is only with optical biometry devices leading to almost 100% emmetropia.72

Author Contributions

All authors made a significant contribution to the work reported, whether that is in the conception, study design, execution, acquisition of data, analysis, and interpretation, or in all these areas; took part in drafting, revising, or critically reviewing the article; gave final approval of the version to be published; have agreed on the journal to which the article has been submitted; and agree to be accountable for all aspects of the work.

Funding

No external support, either public or private, was received for the conduct of this study.

Disclosure

The authors declare that they have no competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

References

1. Astbury N, Ramamurthy B. How to avoid mistakes in biometry. Community Eye Health. 2006;19(60):70–71.

2. Wang B, Naidu RK, Qu X. Factors related to axial length elongation and myopia progression in orthokeratology practice. PLoS One. 2017;12(4):e0175913.

3. Kanclerz P, Khoramnia R, Wang X. Current developments in corneal topography and tomography. Diagnostics. 2021;11(8):1466. doi:10.3390/diagnostics11081466

4. Motlagh MN, Moshirfar M, Murri MS, et al. Pentacam® corneal tomography for screening of refractive surgery candidates: a review of the literature, part I. Med Hypothesis Discov Innov Ophthalmol. 2019;8(3):177–203.

5. Flaxman SR, Bourne RRA, Resnikoff S, et al. Global causes of blindness and distance vision impairment 1990-2020: a systematic review and meta-analysis. Lancet Glob Health. 2017;5(12):e1221–e1234. doi:10.1016/S2214-109X(17)30393-5

6. Findl O, Drexler W, Menapace R, Heinzl H, Hitzenberger CK, Fercher AF. Improved prediction of intraocular lens power using partial coherence interferometry. J Cataract Refract Surg. 2001;27(6):861–867. doi:10.1016/S0886-3350(00)00699-4

7. Kolega MŠ, Kovačević S, Čanović S, Pavičić AD, Bašić JK. Comparison of IOL--master and ultrasound biometry in preoperative intra ocular lens (IOL) power calculation. Coll Antropol. 2015;39(1):233–235.

8. Kunert KS, Peter M, Blum M, et al. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42(1):76–83. doi:10.1016/j.jcrs.2015.07.039

9. Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 1992;18(2):125–129. doi:10.1016/S0886-3350(13)80917-0

10. Holladay JT. Ultrasound and optical biometry. Cata Refra Surg Global Eur. 2009;4::18‑19.

11. Németh J, Fekete O, Pesztenlehrer N. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg. 2003;29(1):85–88. doi:10.1016/S0886-3350(02)01500-6

12. Findl O. Biometry and intraocular lens power calculation. Curr Opin Ophthalmol. 2005;16(1):61–64. doi:10.1097/00055735-200502000-00011

13. Fercher AF, Mengedoht K, Werner W. Eye-length measurement by interferometry with partially coherent light. Opt Lett. 1988;13(3):186–188. doi:10.1364/OL.13.000186

14. Vogel A, Dick HB, Krummenauer F. Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliability. J Cataract Refract Surg. 2001;27(12):1961–1968. doi:10.1016/s0886-3350(01)01214-7

15. Hoffer KJ, Shammas HJ, Savini G. Comparison of 2 laser instruments for measuring axial length [published correction appears in J cataract refract surg. 2010 6):1066]. J Cataract Refract Surg. 2010;36(4):644–648. doi:10.1016/j.jcrs.2009.11.007

16. Mandal P, Berrow EJ, Naroo SA, et al. Validity and repeatability of the aladdin ocular biometer [published correction appears in br j ophthalmol. 2015 dec; 99(12):1746]. Br J Ophthalmol. 2014;98(2):256–258. doi:10.1136/bjophthalmol-2013-304002

17. Huang J, Savini G, Li J, et al. Evaluation of a new optical biometry device for measurements of ocular components and its comparison with IOLMaster. Br J Ophthalmol. 2014;98(9):1277–1281. doi:10.1136/bjophthalmol-2014-305150

18. Srivannaboon S, Chirapapaisan C, Chonpimai P, Koodkaew S. Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometer. J Cataract Refract Surg. 2014;40(5):709–715. doi:10.1016/j.jcrs.2013.09.020

19. Ventura BV, Ventura MC, Wang L, Koch DD, Weikert MP. Comparison of biometry and intraocular lens power calculation performed by a new optical biometry device and a reference biometer. J Cataract Refract Surg. 2017;43(1):74–79. doi:10.1016/j.jcrs.2016.11.033

20. Jung S, Chin HS, Kim NR, Lee KW, Jung JW. Comparison of repeatability and agreement between swept-source optical biometry and dual-scheimpflug topography. J Ophtha. 2017;2017:1516395. doi:10.1155/2017/1516395

21. Huang J, Savini G, Hoffer KJ, et al. Repeatability and interobserver reproducibility of a new optical biometer based on swept-source optical coherence tomography and comparison with IOLMaster. Br J Ophthalmol. 2017;101(4):493–498. doi:10.1136/bjophthalmol-2016-308352

22. Goebels S, Pattmöller M, Eppig T, Cayless A, Seitz B, Langenbucher A. Comparison of 3 biometry devices in cataract patients. J Cataract Refract Surg. 2015;41(11):2387–2393. doi:10.1016/j.jcrs.2015.05.028

23. Jiang J, Pan X, Zhou M, Wang X, Zhu H, Li D. A comparison of IOLMaster 500 and IOLMaster 700 in the measurement of ocular biometric parameters in cataract patients. Sci Rep. 2022;12(1):12770. doi:10.1038/s41598-022-16985-8

24. Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238(9):765–773. doi:10.1007/s004170000188

25. Bullimore MA, Buehren T, Bissmann W. Agreement between a partial coherence interferometer and 2 manual keratometers. J Cataract Refract Surg. 2013;39(10):1550–1560. doi:10.1016/j.jcrs.2013.03.034

26. Skrzypecki J, Grabska-Liberek I, Guszkowska M, Izdebska J, Szaflik JP. Immersion biometry for intraocular lens power calculation with fourth-generation formulas. Clin Ophthalmol. 2020;14:2159–2162. doi:10.2147/OPTH.S259078

27. Xia T, Martinez CE, Tsai LM. Update on intraocular lens formulas and calculations. Asia Pac J Ophthalmol. 2020;9(3):186–193. doi:10.1097/APO.0000000000000293

28. Cione F, La Padula G, Formicola A. Axial length correction in evaluation of refractive predictability and biometry agreement [Letter]. Clin Ophthalmol. 2023;17:2357–2358. doi:10.2147/OPTH.S433664

29. Chen YA, Hirnschall N, Findl O. Evaluation of 2 new optical biometry devices and comparison with the current gold standard biometer. J Cataract Refract Surg. 2011;37(3):513–517. doi:10.1016/j.jcrs.2010.10.041

30. Cruysberg LP, Doors M, Verbakel F, Berendschot TT, De Brabander J, Nuijts RM. Evaluation of the lenstar LS 900 non-contact biometer. Br J Ophthalmol. 2010;94(1):106–110. doi:10.1136/bjo.2009.161729

31. McAlinden C, Wang Q, Pesudovs K, et al. Axial length measurement failure rates with the IOLMaster and lenstar ls 900 in eyes with cataract. PLoS One. 2015;10(6):e0128929. doi:10.1371/journal.pone.0128929

32. Cione F, Pellegrino I, Di Paola I. Comment on: efficacy of segmented axial length and artificial intelligence approaches to intraocular lens power calculation in short eyes. J Cataract Refract Surg. 2023;49(11):1187. doi:10.1097/j.jcrs.0000000000001284

33. Hoffer KJ, Savini G. Update on intraocular lens power calculation study protocols: the better way to design and report clinical trials. Ophthalmology. 2021;128(11):e115–e120. doi:10.1016/j.ophtha.2020.07.005

34. Reitblat O, Levy A, Kleinmann G, Assia EI. Accuracy of intraocular lens power calculation using three optical biometry measurement devices: the OA-2000, Lenstar-LS900 and IOLMaster-500. Eye. 2018;32(7):1244–1252. doi:10.1038/s41433-018-0063-x

35. Ha A, Wee WR, Kim MK. Comparative efficacy of the new optical biometer on intraocular lens power calculation (AL-Scan versus IOLMaster). Korean J Ophthalmol. 2018;32(3):241–248. doi:10.3341/kjo.2017.0063

36. Huang J, Savini G, Wu F, et al. Repeatability and reproducibility of ocular biometry using a new noncontact optical low-coherence interferometer. J Cataract Refract Surg. 2015;41(10):2233–2241. doi:10.1016/j.jcrs.2015.10.062

37. Kaswin G, Rousseau A, Mgarrech M, Barreau E, Labetoulle M. Biometry and intraocular lens power calculation results with a new optical biometry device: comparison with the gold standard. J Cataract Refract Surg. 2014;40(4):593–600. doi:10.1016/j.jcrs.2013.09.015

38. Shi Q, Wang GY, Cheng YH, Pei C. Comparison of IOL-master 700 and IOL-master 500 biometers in ocular biological parameters of adolescents. Int J Ophthalmol. 2021;14(7):1013–1017.

39. Guimarães de Souza R, Montes de Oca I, Esquenazi I, Al-Mohtaseb Z, Weikert MP. Updates in biometry. Int Ophthalmol Clin. 2017;57(3):115–124. doi:10.1097/IIO.0000000000000175

40. Akman A, Asena L, Güngör SG. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol. 2016;100(9):1201–1205. doi:10.1136/bjophthalmol-2015-307779

41. Lawless M, Jiang JY, Hodge C, Sutton G, Roberts TV, Barrett G. Total keratometry in intraocular lens power calculations in eyes with previous laser refractive surgery. Clin Exp Ophthalmol. 2020;48(6):749–756. doi:10.1111/ceo.13760

42. De Bernardo M, Cione F, Capasso L, Coppola A, Rosa N. A formula to improve the reliability of optical axial length measurement in IOL power calculation. Sci Rep. 2022;12(1):18845.

43. Omoto MK, Torii H, Masui S, Ayaki M, Tsubota K, Negishi K. Ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers with segmental or equivalent refractive indices [published correction appears in sci rep. 10(1):13181]. Sci Rep. 2019;9(1):6557. doi:10.1038/s41598-019-42968-3

44. Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophtha Sci Res. 2017;55(3):1.

45. Shammas HJ, Ortiz S, Shammas MC, Kim SH, Chong C. Biometry measurements using a new large-coherence-length swept-source optical coherence tomographer. J Cataract Refract Surg. 2016;42(1):50–61. doi:10.1016/j.jcrs.2015.07.042

46. Montés-Micó R. Evaluation of 6 biometers based on different optical technologies. J Cataract Refract Surg. 2022;48(1):16–25. doi:10.1097/j.jcrs.0000000000000690

47. Hoffer KJ, Shammas HJ, Savini G, Huang J. Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and China. J Cataract Refract Surg. 2016;42(1):62–67. doi:10.1016/j.jcrs.2015.07.041

48. Xiong Y, Lin Y, Zhao Z, Wang H, Zhang G. Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes. BMC Ophthalmol. 2023;23(1):218. doi:10.1186/s12886-023-02962-w

49. Savini G, Hoffer KJ, Shammas HJ, Aramberri J, Huang J, Barboni P. Accuracy of a new swept-source optical coherence tomography biometer for IOL power calculation and comparison to IOLMaster. J Refract Surg. 2017;33(10):690–695. doi:10.3928/1081597X-20170721-05

50. Sikorski BL, Suchon P. OCT biometry (B-OCT): a new method for measuring ocular axial dimensions. J Ophtha. 2019;2019:1–10. doi:10.1155/2019/9192456

51. Wylęgała A, Bolek B, Mazur R, Wylęgała E. Repeatability, reproducibility, and comparison of ocular biometry using a new optical coherence tomography-based system and another device. Sci Rep. 2020;10(1):14440.

52. Wylęgała A, Mazur R, Bolek B, Wylęgała E. Reproducibility, and repeatability of corneal topography measured by revo NX, Galilei g6 and casia 2 in normal eyes. PLoS One. 2020;15(4):e0230589. doi:10.1371/journal.pone.0230589

53. Moshirfar M, Tenney S, McCabe S, Schmid G. Repeatability and reproducibility of the Galilei G6 and its agreement with the pentacam® AXL in optical biometry and corneal tomography. Expert Rev Med Devices. 2022;19(4):375–383. doi:10.1080/17434440.2022.2075725

54. Ladi JS, Shah NA. Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry. Indian J Ophthalmol. 2010;58(5):385–388. doi:10.4103/0301-4738.67045

55. Cione F, De Bernardo M, Gioia M, et al. A no-history multi-formula approach to improve the iol power calculation after laser refractive surgery: preliminary results. J Clin Med. 2023;12(8):2890. doi:10.3390/jcm12082890

56. Moshirfar M, Buckner B, Ronquillo YC, Hofstedt D. Biometry in cataract surgery: a review of the current literature. Curr Opin Ophthalmol. 2019;30(1):9–12. doi:10.1097/ICU.0000000000000536

57. Kapoor M, Venkatesh P, Chawla R, Temkar S, Aggarwal E. Simplifying biometry in oil-filled eyes: a novel formula for axial length calculation in eyes with 1000 cSt silicone oil. Indian J Ophthalmol. 2023;71(6):2466–2468. doi:10.4103/ijo.IJO_2187_22

58. Kane JX, LaHood BR, Goggin M. Analysis of posterior corneal surgically induced astigmatism following cataract surgery with a 1.8-mm temporal clear corneal incision. J Refract Surg. 2023;39(6):381–386. doi:10.3928/1081597X-20230426-01

59. Bao T, Wang L, Liu C, Yang Y, Pang Y. Analysis of biometric parameters of cataract eyes measured with optical biometer Lenstar LS900, IOL Master 700, and OPD-SCAN III. Photodiagnosis Photodyn Ther. 2023;43:103646. doi:10.1016/j.pdpdt.2023.103646

60. Röggla V, Langenbucher A, Leydolt C, et al. Best fit formula approach in delayed sequential bilateral cataract surgery. Clin Exp Ophthalmol. 2023;51(6):559–565. doi:10.1111/ceo.14261

61. Domínguez-Vicent A, Venkataraman AP, Dalin A, Brautaset R, Montés-Micó R. Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer. Eye Vis. 2023;10(1):24.

62. Gjerdrum B, Gundersen KG, Nilsen C, Gundersen M, Jensen P. Refractive predictability and biometry agreement of a combined swept source optical coherence and reflectometry biometer compared to an optical low coherence reflectometry biometer and an SS-OCT Biometer. Clin Ophthalmol. 2023;17:1439–1452.

63. Nihalani BR, Oke I, VanderVeen DK. Comparison of baseline biometry measures in eyes with pediatric cataract to age-matched controls. Graefes Arch Clin Exp Ophthalmol. 2023;261(10):3007–3013. doi:10.1007/s00417-023-06122-0

64. Neoh FP, A Y, Siti AA, Liza-Sharmini AT. Anterior Segment Biometry In Primary Angle Closure Glaucoma Patients With Visual Field Progression: Comparison Between Malays and Chinese. J Curr Glaucoma Pract. 2023;17(1):3–8. doi:10.5005/jp-journals-10078-1391

65. Zhao MH, Song Y, Liu JL, et al. Investigation of ocular biometry in 4- to 9-year-old Chinese children. BMC Ophthalmol. 2023;23(1):225. doi:10.1186/s12886-023-02975-5

66. Badakere A, Ghaisas SP, Akshya P, Natarajan V, Guha S, Agarkar S. Intraocular lens formula calculation in pediatric eyes: do we have an answer? A retrospective comparison between sanders-retzlaff-kraff ii and barret’s formula. Indian J Ophthalmol. 2023;71(5):2139–2142. doi:10.4103/IJO.IJO_3191_22

67. Mukhija R, Vanathi M, Verma M, Raj N, Gupta N, Tandon R. Comparative evaluation of intraoperative aberrometry and Barrett’s toric calculator in toric intraocular lens implantation. Indian J Ophthalmol. 2023;71(5):1918–1923. doi:10.4103/ijo.IJO_2092_22

68. Sivakumar T, Palmowski-Wolfe A. Retrospective comparison of the myopia master and the lenstar ls900 axial length measurements in children with myopia [published correction appears in klin monbl augenheilkd. 2023 apr;240(4):e2]. retrospektiver vergleich der axiallängenmessungen mit myopia master und lenstar ls900 bei kindern mit myopie [published correction appears in klin monbl augenheilkd. 2023 apr; 240(4):e2]. Klin Monbl Augenheilkd. 2023;240(4):587–590. doi:10.1055/a-2013-2557

69. Tañá-Rivero P, Tello-Elordi C, Orts-Vila P, Tañá-Sanz P, Tañá-Sanz S. Agreement of corneal diameter and anterior chamber depth measurements with the IOLMaster 500 and the IOLMaster 700 optical biometers in myopic eyes. Clin Ophthalmol. 2023;17:1245–1253. doi:10.2147/OPTH.S409132

70. Michael R, Wirkner K, Engel C, Loeffler M, Kirsten T, Rauscher FG. Feasibility and repeatability of ocular biometry measured with IOLMaster 700 in a large population-based study. Ophthalmic Physiol Opt. 2023;43(4):860–873. doi:10.1111/opo.13148

71. Lin L, Xu M, Mo E, et al. Accuracy of newer generation iol power calculation formulas in eyes with high axial myopia. J Refract Surg. 2021;37(11):754–758. doi:10.3928/1081597X-20210712-08

72. Tañá-Rivero P, Tañá-Sanz S, Pastor-Pascual F, Ruiz-Mesa R, Montés-Micó R. Axial length measurement failure rates using optical biometry based on swept-source OCT in cataractous eyes. Expert Rev Med Devices. 2022;19(8):633–640. doi:10.1080/17434440.2022.2118047

73. Kongsap P. Comparison of a new optical biometer and a standard biometer in cataract patients. Eye Vis. 2016;3(1):27. doi:10.1186/s40662-016-0059-1

Creative Commons License © 2024 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.