Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique

Authors Kanellopoulos J , Asimellis G

Received 28 October 2012

Accepted for publication 27 November 2012

Published 15 February 2013 Volume 2013:7 Pages 343—351

DOI https://doi.org/10.2147/OPTH.S39644

Review by Single anonymous peer review

Peer reviewer comments 3



Video abstract presented by John Kanellopoulos

Views: 300

A John Kanellopoulos,1,2 George Asimellis1

1Laservision Eye Institute, Athens, Greece; 2New York University School of Medicine, New York, NY, USA

Background: The purpose of this study was to determine flap parameter accuracy, extent of the opaque bubble layer, and incidence of skip lines in femtosecond laser-assisted stromal in situ keratomileusis (LASIK) using the WaveLight® FS200 laser and optoelectronic clinical measurements.
Methods: Images from 101 flaps were automatically recorded during consecutive routine LASIK procedures performed using the WaveLight FS200 femtosecond laser and the EX500 excimer laser. Digital processing of these images was used to evaluate objectively the diameter of FS200-created flaps, by comparing planned versus achieved procedures and to evaluate the incidence and extent (area) of the opaque bubble layer.
Results: The intended flap diameters were between 8.00 mm and 9.50 mm. The achieved flap diameters showed extremely high precision, and were on average -0.16 ± 0.04 mm smaller for a 8.00 mm intended flap diameter, -0.12 ± 0.03 mm smaller for a 8.50 mm flap, and up +0.06 ± 0.06 mm wider for a 9.50 mm flap. With an average flap area of 72.4 mm2, the mean area of the opaque bubble layer (4.1 ± 4.3 [range 0–14.34] mm2) corresponded to a 6% opaque bubble layer-to-flap area. Specifically, 80% of the femtosecond-created flaps had an essentially zero opaque bubble layer (<2.7% of the flap area).
Conclusion: In our clinical experience, flaps created using FS200 and this novel highly objective assessment technique demonstrate both precision and reproducibility. The incidence of opaque bubble layer was minimal.

Keywords: femtosecond laser precision, bladeless laser-assisted stromal in situ keratomileusis, corneal flap diameter, opaque bubble layer, skip lines, WaveLight FS200

Creative Commons License © 2013 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.