Visual Function After Implantation of Trifocal and Trifocal Toric Intraocular Lenses Using Intraoperative Aberrometry

Authors Brissette A, Cole B, Hall B 

Received 21 November 2023

Accepted for publication 28 April 2024

Published 30 May 2024 Volume 2024:18 Pages 1547—1554

DOI https://doi.org/10.2147/OPTH.S450979

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 3

Editor who approved publication: Dr Scott Fraser



Ashley Brissette,1 Brigette Cole,1 Brad Hall2

1Weill Cornell Medicine, New York Presbyterian Hospital, New York, NY, USA; 2Sengi, Penniac, NB, Canada

Correspondence: Ashley Brissette, Weill Cornell Medicine, New York Presbyterian Hospital, 1305 York Ave, New York, NY, 10021, USA, Tel +1 646-962-2020, Email [email protected]

Purpose: To evaluate patient outcomes and visual function following trifocal and trifocal toric intraocular lens (IOL) implantation using intraoperative aberrometry at a single site in the US.
Methods: This prospective, single arm study included 21 subjects that completed 3 month follow-up. Inclusion criteria were visually significant cataract and potential post-operative visual acuity of 20/25 or better. Endpoints included postoperative prediction error, refractive outcomes, uncorrected visual acuities at distance (UDVA), intermediate (UIVA), and near (UNVA), contrast sensitivity, and subject responses on the modified Visual Function Quality of Life Questionnaire (VF-14 QOL).
Results: Binocular UDVA, UIVA, and UNVA were 20/25 or better in 100% (21/21), 100% (21/21), 90% (19/21) of subjects. The absolute prediction error was 0.50 D or less in 79% (33/42) of eyes, and 81% (34/42) and 86% (36/42) of eyes achieved ≤ 0.5 D of residual astigmatism and manifest refraction spherical equivalent, respectively. On the modified VF-14 QOL, driving at night, reading small print, and reading a newspaper or book were the tasks that had the lowest percentages of subjects reporting no difficulty or a little difficulty.
Conclusion: Implantation with trifocal and trifocal toric IOLs using intraoperative aberrometry can provide high refractive precision, leading to excellent visual performance and low visual task difficulty at all ranges (distance, intermediate, and near).

Plain Language Summary: An intraocular lens (IOL) is a clear artificial lens that can be used to replace the natural lens in the eye when the natural lens becomes opaque (develops a cataract). Monofocal IOLs are designed to provide good vision to see distant objects; however, spectacles may still be needed to see objects clearly up close (such as reading a book or using a digital device). Trifocal IOLs are designed to provide good vision to see objects at distance and up close, however, the power of the IOL must be accurately determined for the best visual outcomes.
Devices called biometers are used by cataract surgeons to measure the eye and determine the most appropriate lens power to implant. Most biometers are used prior to surgery, however one type, intraoperative aberrometry (IA), can be used during surgery to measure the eye and determine the most appropriate lens power. The purpose of this study was to evaluate patient outcomes and visual function following trifocal IOL implantation using IA. The results of this study suggest that implantation with trifocal IOLs using IA can provide high refractive accuracy and excellent visual outcomes.

Keywords: trifocal IOL, PanOptix, cataract surgery, intraoperative aberrometry

Introduction

Cataract surgery patients—and surgeons—have high expectation of good clinical outcomes following intraocular lens (IOL) implantation. Postoperative visual outcomes are heavily reliant on achieving the postoperative target refraction especially for trifocal and other advanced technology lenses. Recent studies estimate that 73–80% of eyes are within 0.5 D of target refraction following cataract surgery with modern day biometry and formulas.1,2

Trifocal IOLs are designed to split incoming light and redirect the light into 3 distinct foci, to achieve good visual outcomes at distance intermediate, and near. The first trifocal IOL that was approved by the US FDA was the AcrySof PanOptix IOL (Alcon Vision, LLC). With the PanOptix lens, 50% of incoming light is distributed for distance vision, 25% intermediate, and 25% for near vision.3 The PanOptix lens has been reported to provide good visual outcomes.4–6 However, as with other multifocal or presbyopia correcting IOLs, visual disturbances have been reported.7–10

As lens technology advances, so do patient expectations for excellent refractive outcomes after cataract surgery. The goal of the modern day surgeon is to improve the percentage of eyes within targeted refraction and thereby improve patient satisfaction and decreased dependence on glasses and contact lenses after surgery. Biometry with advanced lens formulas have proven effective at improving our lens calculations, however refractive surprises still exist, especially for more complex eyes like extremes of axial length. It is especially crucial for optimal performance of trifocal lenses to achieve the postoperative target refraction due to their complex optics. Missing the refractive target may significantly impact patient postoperative visual function and satisfaction. Intraoperative aberrometry (IA) is a tool that allows precise measurements of aphakic and pseudophakic eyes during cataract surgery. Using IA in aphakic eyes can allow for IOL power determination.11 Using IA in pseudophakic eyes can assist in determining if any adjustments to the lens position are needed and is especially helpful in toric IOLs.10 The only currently approved IA device is the ORA system (Alcon Vision, LLC). The use of ORA has been reported to improve refractive outcomes compared to preoperative calculations alone.12–15 Though there have been reports of no differences in prediction error between ORA and preoperative planning.16,17

Many reports of outcomes with IA have focused on toric IOLs.13,18,19 With the increasing popularity of multifocal IOLs, it is important to understand the predictive accuracy and refractive outcomes with IA in this patient cohort. There have been a few reports on refractive outcomes with intraoperative aberrometry and trifocal IOLs outside the US,12,20,21 however, there is little data regarding outcomes within the US. In addition, many of the published OUS studies were retrospective. The purpose of this study is to prospectively evaluate patient outcomes and visual function following trifocal IOL implantation using intraoperative aberrometry at a single site in the US.

Methods

This was a single arm, prospective study of visual function following bilateral cataract surgery and implantation with a trifocal IOL. This study was reviewed and approved by an institutional review board (IRB, Weill Cornell Medicine). Subjects gave written informed consent before participation in the study. This study was conducted in compliance with Good Clinical Practice (GCP), International Harmonization (ICH) guidelines, and the tenets of the Declaration of Helsinki. This study was registered on clinicaltrials.gov (NCT04196569). Data are not available for sharing.

Subjects were included if they had visually significant cataract and potential post-operative visual acuity of 20/25 or better. Exclusion criteria were irregular astigmatism, prior corneal refractive surgery, or ocular pathology contraindicated for trifocal IOLs (including but not limited to diabetic retinopathy, age-related macular degeneration or other macular pathology, cornea guttata, corneal scarring, corneal ectasia, glaucoma with visual field loss).

Preoperative biometry measurements were obtained using Veracity and IOL Master 700 (Carl Zeiss Medica) with Barrett Total Keratometry Formula to determine IOL power. One experienced surgeon performed all surgeries (AB). The dispersive viscoelastic was removed from the eye, and cohesive viscoelastic was instilled in a uniform manner. Intraocular pressure was measured with the Ocular Kasaby Barraquer Tonometer to validate fill. All patients were offered options for Femtosecond Laser Assisted Cataract Surgery (FLACS) and manual cataract surgery. If they opted for FLACS, capsulotomy, lens fragmentation, and limbal relaxing incision (if required) were performed using a LenSx femtosecond laser (Alcon Vision, LLC). The VERION System was used for image-guided digital tracking. The ORA System with Verifeye+ (Alcon Vision, LLC) was used intraoperatively to determine IOL power, cylinder power, and final axis of placement. Where the powers suggested by preoperative biometry and intraoperative aberrometry differed, the power suggested by intraoperative aberrometry was selected. AcrySof IQ PanOptix non-toric and toric IOL models, were implanted (TFAT00, TFAT30, TFAT40; Alcon Vision, LLC). The postoperative regimen was the surgeon’s preferred standard of care.

Study endpoints included the distribution of residual refractive error at 3 months postoperative, uncorrected visual acuities at distance (UDVA; 4m), intermediate (UIVA; 60cm), and near (UNVA; 40cm) at 1 and 3 months postoperative, contrast sensitivity testing at 1 and 3 months postoperative, and subject responses on the modified Visual Function Quality of Life Questionnaire (VF-14 QOL) at 1 and 3 months postoperative.22 Contrast sensitivity was measured using a scale from 100% to 0.8% with 100% being highest contrast and 0.8% being lowest contrast. Subjects were shown 2 letters at the 20/100 line for each contrast level and worked down the scale until they reached the lowest contrast that they could see.

This study was intended to be descriptive, thus there was no a priori sample size calculation. However, it is worth noting that sample sizes of 20 to 30 have been successfully used to characterize the performance of trifocal IOLs.4,20 The target sample size for this study was 25 subjects. Statistical analyses were performed using the software program R (version 4.2.2; The R Foundation for Statistical Computing, Vienna, Austria).

Results

A total of 25 subjects (50 eyes) were enrolled in this study. Three patients did not complete study due to COVID restrictions at the time. One subject did not complete the Month 3 visit as they were lost to follow up. The preoperative and patient demographics are summarized in Table 1. There were no adverse events related to the IOL or surgery.

Table 1 Preoperative Patient Demographics

The postoperative refractive outcomes are summarized in Table 2. Refractive outcomes were excellent, with 81% (34/42) and 86% (36/42) of eyes achieving ≤0.5 D of residual astigmatism and manifest refraction spherical equivalent, respectively. In addition, 79% (33/42) of eyes had ≤0.5 D of absolute prediction error using IA. The powers suggested by IA and preoperative planning were different in 24% (10/42) of eyes. All suggested power differences were within 0.5 D. Mean axial length in the cases where the powers differed was 25.61 ± 1.12 mm compared to 24.60 ± 1.92 mm in the cases where the suggested powers were the same. The differences in mean axial lengths between these cases was statistically significant using the Welch two sample t-test (p = 0.048).

Table 2 Postoperative Refractive Outcomes (n = 42 Eyes)

Monocular and binocular visual acuities are summarized in Table 3 and Figure 1. Monocular visual acuities at postoperative month 3 were generally higher than at postoperative month 1, with 93% (39/42), 100% (42/42), 98% (41/42) of eyes 20/32 or better at distance, intermediate, and near. Binocular visual acuities at postoperative month 3 were also generally higher than at postoperative month 1, with 100% (21/21), 100% (21/21), 90% (19/21) of patients 20/25 or better at distance, intermediate, and near.

Table 3 Postoperative Visual Outcomes

Figure 1 Monocular uncorrected visual acuities at (A) Distance, (B) Intermediate, and (C) Near and binocular uncorrected visual acuities at (D) Distance, (E) Intermediate, and (F) Near.

Table 4 summarizes the percentage of subjects reporting “No Difficulty” or “A Little Difficulty” on the Visual Function Quality of Life Questionnaire. Visual tasks are grouped by distance, intermediate, and near in Table 4. Difficulty for distance visual tasks was low, with greater than 80% of subjects reporting none or little difficulty at postoperative months 1 and 3 for all tasks, with the exception of the task Driving at night. Difficulty for intermediate visual tasks was also low, with greater than 85% of subjects reporting none or little difficulty at postoperative months 1 and 3 for all tasks. Difficulty was higher for near tasks, compared to intermediate and distance tasks. However, difficulty was still acceptable with 75% or more subjects reporting none or little difficulty at postoperative month 3 for all tasks.

Table 4 Postoperative Task “No Difficulty” or “A Little Difficulty” (Percentage of Subjects)

Table 5 summarizes the postoperative month 1 and 3 monocular and binocular contrast sensitivity. Contrast sensitivity was good, with all patients able to see at 4% contrast. As expected, binocular contrast sensitivity was higher than monocular contrast sensitivity. The differences between postoperative months 1 and 3 were minor.

Table 5 Postoperative Monocular (n = 42 Eyes) and Binocular (n = 19 Subjects) Contrast Sensitivity

Discussion

Good refractive outcomes following cataract surgery and IOL implantation are crucial for good visual outcomes. In this study, we evaluated the refractive outcomes, visual outcomes, and visual function of subjects who had IA performed intraoperatively and received a trifocal IOL bilaterally. In general, the results of our study suggest that clinical outcomes are excellent using IA and trifocal IOLs.

The prediction error with IA and trifocal implantation was good, with 79% of eyes within ± 0.5 D of the predicted refraction. Ma et al20 observed similar results to ours, with 82% of eyes within ± 0.5 D of the predicted refraction using IA with trifocal implantation. Blaylock and Hall12 reported APE ≤ 0.5 D in 88% of eyes using IA with trifocal implantation, higher than our study. In addition, Watanabe21 observed ≤0.5 D in 93% of eyes (after optimization) using IA with trifocal implantation. Differences between our study and previous reports of IA with trifocals12,20,21 may be explained by the lower sample size in our study, the retrospective design of the other studies, or patient populations.

The manifest refraction spherical equivalent (MRSE) ≤0.5 D was achieved in 86% of study eyes. This is a similar result to the United States Food and Drug Administration trial,23 which reported MRSE ≤ 0.5 D was achieved in 84% of study eyes with the PanOptix (non-toric only). Scheepers et al24 observed that MRSE ≤ 0.5 D was achieved in 84% of study eyes with the PanOptix. In a retrospective study, Carreno et al25 reported MRSE ≤ 0.5 D in 96% of study eyes using the PanOptix. In another retrospective study, Lawless et al26 observed MRSE ≤ 0.5 D in 100% of study eyes with the PanOptix. Differences between our study and previous reports of the PanOptix may be explained by the lower sample size in our study or patient populations. Based on the results of our study and previous reports, it appears that implantation of a trifocal using IA can results in excellent refractive outcomes.

Visual outcomes were excellent in our study, with binocular UDVA, UIVA, and UNVA 20/25 or better in 100%, 100%, 90% of patients. Kohnen et al,5 reported on the visual outcomes from 547 bilaterally implanted patients with the TFNT00 IOL, with binocular UDVA, UIVA, and UNVA 20/25 or better in 92%, 87%, 88% of patients, similar to our study. Other studies have also reported good visual outcomes with the PanOptix trifocal IOL at distance, intermediate and near.4,6,10,12,23,24 The results of our study and other suggest that the PanOptix trifocal IOL using IA can provide good visual outcomes.

Visual function and quality of life were evaluated using a questionnaire. We used the modified VF-14 QOL questionnaire reported by Akman et al.22 In their study, the authors observed generally low patient report difficulty with performing tasks, though driving at night, reading small print, and doing fine hand work were the tasks with the highest patient reported difficulty. The results from our study are similar, with driving at night, reading small print, and reading a newspaper or book the tasks that had the lowest percentages of subjects reporting no difficulty or a little difficulty. It should be noted that in our study we are reporting the results of each question on the questionnaire as percentages of respondents, as we assume that the questionnaire scales are ordinal not interval.27

A limitation of this study was the relatively short follow up period. Three months postoperative is a typical time frame to characterize clinical outcomes; however, we are not able to draw any conclusions about the long-term clinical outcomes. Another limitation was the relatively small sample size as the first few patients were enrolled right before the COVID-19 pandemic and were subsequently lost to follow up due to New York State’s strict regulations for elective surgery at the time. However, a final sample size of 22 subject has shown to be sufficient to describe the performance of IA and the trifocal IOL.

In conclusion, the results of this study suggest that implantation with trifocal and trifocal toric IOLs using intraoperative aberrometry can provide high refractive precision, leading to excellent visual performance and low visual task difficulty at all ranges (distance, intermediate, and near).

Data Sharing Statement

Data are not available for sharing.

Acknowledgment

This paper was presented at the 2023 American Society of Cataract and Refractive Surgery (ASCRS) Annual Meeting as a conference talk with interim findings.

Funding

This study was supported with an investigator-initiated study grant (50746627) from Alcon Vision LLC, Fort Worth, TX, USA.

Disclosure

Dr Brissette is a consultant for Alcon and Carl Zeiss Medica. Brad Hall reports that he has received consulting fees from Ace Vision Group outside the submitted work. The authors report no other conflict of interest for this work.

References

1. Lundstrom M, Dickman M, Henry Y, et al. Risk factors for refractive error after cataract surgery: analysis of 282 811 cataract extractions reported to the European Registry of Quality Outcomes for cataract and refractive surgery. J Cataract Refract Surg. 2018;44(4):447–452. doi:10.1016/j.jcrs.2018.01.031

2. Melles RB, Holladay JT, Chang WJ. Accuracy of intraocular lens calculation formulas. Ophthalmology. 2018;125(2):169–178. doi:10.1016/j.ophtha.2017.08.027

3. Garcia-Perez JL, Gros-Otero J, Sanchez-Ramos C, Blazquez V, Contreras I. Short term visual outcomes of a new trifocal intraocular lens. BMC Ophthalmol. 2017;17:72.

4. Blehm C, Potvin R. Reported patient satisfaction and spectacle independence following bilateral implantation of the PanOptix((R)) trifocal intraocular lens. Clin Ophthalmol. 2021;15:2907–2912. doi:10.2147/OPTH.S323337

5. Kohnen T, Lapid-Gortzak R, Ramamurthy D, et al. Clinical outcomes after bilateral implantation of a diffractive trifocal intraocular lens: a worldwide pooled analysis of prospective clinical investigations. Clin Ophthalmol. 2023;17:155–163. doi:10.2147/OPTH.S377234

6. Kohnen T, Marchini G, Alfonso JF, et al. Innovative trifocal (quadrifocal) presbyopia-correcting IOLs: 1-year outcomes from an international multicenter study. J Cataract Refract Surg. 2020;46(8):1142–1148. doi:10.1097/j.jcrs.0000000000000232

7. Cao K, Friedman DS, Jin S, et al. Multifocal versus monofocal intraocular lenses for age-related cataract patients: a system review and meta-analysis based on randomized controlled trials. Surv Ophthalmol. 2019;64(5):647–658. doi:10.1016/j.survophthal.2019.02.012

8. Hovanesian JA, Jones M, Allen Q. The PanOptix trifocal IOL vs the ReSTOR 2.5 active focus and ReSTOR 3.0-add multifocal lenses: a study of patient satisfaction, visual disturbances, and uncorrected visual performance. Clin Ophthalmol. 2021;15:983–990. doi:10.2147/OPTH.S285628

9. Pedrotti E, Carones F, Talli P, et al. Comparative analysis of objective and subjective outcomes of two different intraocular lenses: trifocal and extended range of vision. BMJ Open Ophthalmol. 2020;5(1):e000497. doi:10.1136/bmjophth-2020-000497

10. Schallhorn JM. Multifocal and extended depth of focus intraocular lenses: a comparison of data from the United States food and drug administration premarket approval trials. J Refract Surg. 2021;37(2):98–104. doi:10.3928/1081597X-20201111-02

11. Hatch KM, Woodcock EC, Talamo JH. Intraocular lens power selection and positioning with and without intraoperative aberrometry. J Refract Surg. 2015;31(4):237–242. doi:10.3928/1081597X-20150319-03

12. Blaylock JF, Hall B. Clinical outcomes of a diffractive trifocal intraocular lens with femtosecond laser, digital tracking, and intraoperative aberrometry. Can J Ophthalmol. 2021;57(5):291–296. doi:10.1016/j.jcjo.2021.05.014

13. Blaylock JF, Hall BJ. Clinical outcomes of monofocal toric IOLs using digital tracking and intraoperative aberrometry. Clin Ophthalmol. 2021;15:3593–3600. doi:10.2147/OPTH.S322523

14. Cionni RJ, Dimalanta R, Breen M, Hamilton C. A large retrospective database analysis comparing outcomes of intraoperative aberrometry with conventional preoperative planning. J Cataract Refract Surg. 2018;44(10):1230–1235. doi:10.1016/j.jcrs.2018.07.016

15. Ianchulev T, Hoffer KJ, Yoo SH, et al. Intraoperative refractive biometry for predicting intraocular lens power calculation after prior myopic refractive surgery. Ophthalmology. 2014;121(1):56–60. doi:10.1016/j.ophtha.2013.08.041

16. Davison JA, Potvin R. Preoperative measurement vs intraoperative aberrometry for the selection of intraocular lens sphere power in normal eyes. Clin Ophthalmol. 2017;11:923–929. doi:10.2147/OPTH.S135659

17. Raufi N, James C, Kuo A, Vann R. Intraoperative aberrometry vs modern preoperative formulas in predicting intraocular lens power. J Cataract Refract Surg. 2020;46(6):857–861. doi:10.1097/j.jcrs.0000000000000173

18. Modi SS. Clinical outcomes after aphakic versus aphakic/pseudophakic intraoperative aberrometry in cataract surgery with toric IOL implantation. Int Ophthalmol. 2020;40(12):3251–3257. doi:10.1007/s10792-020-01509-y

19. Woodcock MG, Lehmann R, Cionni RJ, Breen M, Scott MC. Intraoperative aberrometry versus standard preoperative biometry and a toric IOL calculator for bilateral toric IOL implantation with a femtosecond laser: one-month results. J Cataract Refract Surg. 2016;42(6):817–825. doi:10.1016/j.jcrs.2016.02.048

20. Ma J, El-Defrawy S, Lloyd J, Rai A. Prediction accuracy of intraoperative aberrometry compared with preoperative biometry formulae for intraocular lens power selection. Can J Ophthalmol. 2021;2:1.

21. Watanabe K. Evaluation of refractive accuracy of ORA and the factors impacting residual astigmatism in patients implanted with trifocal IOLs during cataract surgery: a retrospective observational study. Clin Ophthalmol. 2022;16:2491–2503. doi:10.2147/OPTH.S371555

22. Akman A, Asena L, Ozturk C, Gur Gungor S. Evaluation of quality of life after implantation of a new trifocal intraocular lens. J Cataract Refract Surg. 2019;45(2):130–134.23. doi:10.1016/j.jcrs.2018.12.003

23. Modi S, Lehmann R, Maxwell A, et al. Visual and patient-reported outcomes of a diffractive trifocal intraocular lens compared with those of a monofocal intraocular lens. Ophthalmology. 2020;128(2):197–207. doi:10.1016/j.ophtha.2020.07.015

24. Scheepers MA, Bunce CB, Michaelides M, Hall B. Clinical outcomes of a trifocal compared with an EDOF IOL following bilateral cataract surgery. Can J Ophthalmol. 2022;58(5):393–400. doi:10.1016/j.jcjo.2022.05.005

25. Carreno E, Carreno EA, Carreno R, et al. Refractive and visual outcomes after bilateral implantation of a trifocal intraocular lens in a large population. Clin Ophthalmol. 2020;14:369–376. doi:10.2147/OPTH.S238841

26. Lawless M, Hodge C, Reich J, et al. Visual and refractive outcomes following implantation of a new trifocal intraocular lens. Eye Vis. 2017;4:10.

27. Cohen L, Manion L, Morrison K. Research Methods in Education. Vol. 8. Routledge; 2018.

Creative Commons License © 2024 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.